Heath Consultants EI-5 Ethane Gas Identifier User Manual

Page 10

Advertising
background image

EI-5 Manual Rev A 7

3. Turn on the instrument using the on/off, zeroing switch on the left hand

side of the instrument.

4. Let the instrument warm up for a few minutes. This is indicated by the

graph on the display decreasing towards zero.

5. Zero the instrument using the zero potentiometer on the left side of

instrument.

6. Connect the tubing with the hydrophobic filter in-line to the probe and

the other end to the instrument port labeled Inlet.

7. Switch valve to “Sample” and the pump will draw the sample into the

instrument. Only a few seconds is required.

8. Flip the valve to time and begin timing. The first peak should show on

the graph in approximately 11 to 30 seconds and then approach the

zero point, a second peak (if the sample taken has ethane present)

should appear in 30 to 75 seconds. The second peak will be smaller

and not last as long as the first methane peak.

9. Leave the valve in the “Time” position and let the instrument clear in a

non gaseous environment.

10. Wait 30 seconds after last peak before taking another sample.

11. Remove the hydrocarbon trap and store it back in the case with the

tubing connected at both ends.

12. Turn instrument off.

Hydrocarbon Trap