Heath Consultants EI-5 Ethane Gas Identifier User Manual

Page 6

Advertising
background image

EI-5 Manual Rev A 3

the equilibrium condition in clean air. Any substance foreign to clean air

disturbs this balance, changing the ion flow in the conductance band of

the semi-conductor detects the low concentration and early presence of

foreign gases (methane, ethane, propane) rather than giving a quantita-

tive measure of their concentration. As the various components or hydro-

carbons elute through the separation column they react on the semi-con-

ductor and the resulting change in electrical resistance is amplified and

presented visually on the instrument display.

Flow System: The flow system is so arranged that the Ethane Identifier

cannot be used as a leak locator. It is set up as a field type gas chromato-

graph to differentiate between gases by means of a multi-port selector

valve, short sample loop and separation column.

Valve “Sample” Position: With the selector valve set to “Sample” the

sample is drawn into the instrument and fills the short sample loop. Excess

sample passes through the sample system and exits through the port on

the end of the instrument. The sample does not pass over the sensor in

this position so no meter readings occur.

Valve “Time” Position: After a gas sample has been drawn into the in-

strument, the combustible gas is present in the short sample loop. Switch-

ing the valve to the “Time” position causes the internal pump to push the

sample through the separation column (Chromatograph), over the sensor,

and then exhausts it to the outside of the instrument case at the vent port

located at the bottom of the instrument.

Advertising