TROUBLE SHOOTING

Trouble

Probable Cause

Remedy

No indication on meter

when conducting an

analysis

Instrument being used

in gas-free air

Check another area of

known gas concentra-

tion

Methane / Ethane

peaks overlap and no

down scale display

between peaks

Separation column

breakdown

Sample tested in a

methane background

Gas concentration too

high

Replace separation

column

Connect Hydrocarbon

filter

Take test at location of

lower concentration or

dilute the samples

Methane/ Ethane

peak times are not

correct

Column assembly has

become contaminated

Return instrument to a

Hetek authorized re-

pair facility for column

replacement

Cannot zero

Batteries weak

Replace batteries