Operating instructions – Heath Consultants EI-5 Ethane Gas Identifier User Manual

Page 9

Advertising
background image

EI-5 Manual Rev A 6

OPERATING INSTRUCTIONS

Basic Ethane Identifier Procedure used in a clean background atmo-

sphere:

1. Switch valve to the “Time” position.

2. Turn on the instrument using the on/off, zeroing switch on the left

hand side of the instrument.

3. Let the instrument warm up for a few minutes. This is indicated by the

graph on the display decreasing towards zero.

4. Zero the instrument using the zero potentiometer on the left side of

instrument.

5. Connect the tubing with the hydrophobic filter in-line to the probe and

then connect other end to the instrument port labeled “Inlet”. Take

sample. (10 – 75 LEL works best).

6. Switch valve to “Sample” and the pump will draw the sample into the

instrument. Only a few seconds is required.

7. Flip the valve to time and begin

timing. The first peak should show

on the graph in approximately 10

to 25 seconds and then approach

the zero point, a second peak

(if the sample taken has ethane

present) should appear in 30 to 45

seconds. The second peak will be

smaller and not last as long as the

first methane peak.

8. Leave the valve in the “Time” position and let the instrument clear in a

non gaseous environment.

9. Wait 30 seconds after last peak before taking another sample.

10. Turn instrument off.

Note: The Hetek Ethane Identifier Model EI-5 should not be used on

heavy hydrocarbons. If C4 + hydrocarbons are expected, test for heavy

hydrocarbons by using a charcoal filter with your C.G.I. The Hetek Ethane

Identifier should only be used to differentiate pipeline natural gas from

naturally occurring methane such as marsh gas, swamp gas, landfill gas,

soil gas or methane from other sources.

Ethane Identifier Procedure in a non clean atmosphere:

1. Connect the hydrocarbon trap to the “Carrier” port on the instrument

using the compression fitting. Hand tighten only.

2. Switch valve to the “Time” position.

Advertising