Maintenance instructions – Heath Consultants EI-5 Ethane Gas Identifier User Manual

Page 18

Advertising
background image

EI-5 Manual Rev A 15

Caution: Repair and maintenance beyond that described in this manual

should only be performed by qualified personnel. Return to qualified service

centre for column replacement.

Batteries: If the “Low Battery” LED indicator is solid and bright red, the

batteries require replacing. Four (4) “AA” cells power the instrument. Replace

the batteries by sliding open the battery access panel at the bottom of the

instrument case. Remove old batteries and replace with new batteries.

Observe proper polarity! Check condition of new batteries by observing

the “Low Battery” indicator.

Sensor: The sensor operates at a very low temperature and may never

require replacement. The Figaro 2610 sensor can only be calibrated to a

range of sensitivity and not to a fixed point. It is advisable to return the unit

to an authorized service depot for sensor replacement, as other adjustments

may be necessary to match the sensor to the instrument.

Separation Column: The separation Column included in the instrument

is constructed using Silica Sand. The column should not be disassembled

in any way. It is advisable to return the unit to an authorized service center

for column replacement.

General Maintenance: Keep the instrument clean and preferably stored

in the carrying case supplied with the instrument. As with any instrument,

extremes of temperatures and humidity should be avoided as much as 10

possible. It is not advisable to leave the instrument in an unheated vehicle

night after night in northern climates (winter conditions). Do not subject

the instrument to concentrations of heavy hydrocarbons (hexane, gasoline

fumes, etc.), as the separation column will become contaminated.

MAINTENANCE INSTRUCTIONS

Advertising