Typical elution times – Heath Consultants EI-5 Ethane Gas Identifier User Manual

Page 15

Advertising
background image

EI-5 Manual Rev A 12

Carrier gas regulator set at 6 P.S.I.

Gas

Approximate Time

Methane

10 - 15 Seconds

Ethane

30 - 70 Seconds

Propane

3 - 6 Minutes

Note: The times shown in the above table is the elapsed times between

switching the “Valve” to the “Time” position and the corresponding upscale

indications on the display.

Note: The material used in the separation column has the property of

absorbing and retaining water vapor. Thus the useful life of the separation

column is dependent on the amount of moisture present in the gases being

analyzed.

There may be a need to change the separation column after multiple uses.

The instrument should be returned to a qualified service center for column

replacement.

TYPICAL ELUTION TIMES