Interpreting readings – Heath Consultants EI-5 Ethane Gas Identifier User Manual

Page 14

Advertising
background image

EI-5 Manual Rev A 11

Caution: For accurate indications the temperature and humidity of the

atmosphere being tested should be approximately the same as that of the

instrument.

For optimum results attempt to analyze samples at a concentration of about

10 LEL to 75 LEL. If high concentrations above 10% by volume are analyzed,

proper separation between components cannot be observed. It may be

necessary to take the sample at the fringe area of a suspected pipeline

leak rather than at the highest concentration as established with the C.G.I.

Note: It is important that the operator of the Ethane Identifier take several

tests on their own gas mixture at regular intervals. This will familiarize the

operator with the instrument and will provide an elution (separation) pattern

for that operator’s particular gas. The times listed on instrument might need

to be changed over time as the column ages. A good procedure is to check

on a known gas at least once a month or before use.

INTERPRETING READINGS