15 access records, 1 send records, Access records – Heath Consultants Hi Flow Sampler User Manual

Page 22: Send records

Advertising
background image

Operation

Hi Flow

Sampler

22

Instruction 0055-9017

2. Highlight Manual 1-Stage, and then press the I/O ↵ key.

3. Press ESC to return to the Main Screen. Observe that (M) should appear at the

bottom of the screen.

2.15 Access Records

Note:

Send Records can only be selected if operating in the Expanded Menu Mode. Refer

to Section 2.8.

Selecting Access Records from the Expanded Menu System allows a user to either:

1) send all saved records to a personal computer,

2) view the records on the unit’s LCD screen, or

3) erase all saved records.

2.15.1 Send Records

All the measured and calculated values that were saved in memory can be downloaded

to a personal computer as comma-delimited ASCII text, which can later be imported into

most spreadsheet programs for analysis.

The following describes how to download saved records to a computer that is running

Windows

®

HyperTerminal as the communications program. If a different operating

system and/or communications program is being used, then please consult the

appropriate instruction manuals for those products.

Note:

Send Records can only be selected if operating in the Expanded Menu Mode. Refer

to Section 2.8.

1. Install serial data cable P/N 104-4027 (straight through, 6 foot, DB9 male to DB9

female) between the computer’s COM port and the COMPUTER connector of the Hi
Flow

. (For computers having a 25-pin or USB COM port, the user must supply the

appropriate adapter, or use a serial cable with appropriate connectors on each

end.)

2. On the computer, start HyperTerminal by doing one of the following:

- If HyperTerminal has already been set up to communicate with the Hi Flow

Sampler, then select Start > Programs > Accessories > Communications >

HyperTerminal, and double-click the filename or icon that has been associated

with the instrument.

- If HyperTerminal has never been run in association with the Hi Flow

Sampler,

then a new HyperTerminal connection needs to be made as follows:

a. Select Start > Run. In the Run box type “hypertrm.exe and click OK. The

New Connection dialog box should appear.

b. Type in a New Connection Name such as “HiFlow”. Click OK to select the

default icon (if desired, select a different icon before clicking OK). The

Connect To dialog box should appear.