Hi Flow

Sampler

Operation

Instruction 0055-9017

33

5. The user is now asked to choose a new or different Test ID.

Selecting Yes causes the Access Records menu to appear (refer to Section 2.15).

After selecting or creating a new Test ID, start the measurement by pressing ESC

until the Main Screen appears. Selecting No returns the display to the Main Screen

and starts the measurement using the current Test ID.

6. (Optional) The instrument normally controls the flow rate for measurements #1 and

#2; however, the user can manually set the flow rate using the Speed function. Each
press of the I/O ↵ key will lower the flow rate as indicated by the speed bar moving

to the left.

7. After both measurements are complete, the values in the Main Screen are put on hold

(as indicated by “H--” appearing at the bottom of the screen), allowing the user to make

note of these values before turning off the instrument or proceeding to the next

measurement.

In the examples below, the measured leak concentration was 15% gas during the first

measurement and 18% gas in the second, while the calculated leak rates were 42.5 and

40.8 lpm, respectively. A difference of 4%.

IMPORTANT: For a test to be considered valid, the percent difference between

measurements #1 and #2 must be less than 10%. Scroll up to the #1–#2 function on

the Main Screen to see the percent difference between the last two measurements.

8. When in the automatic mode, measurements are automatically saved. These saved

measurements along with their associated Test IDs can be downloaded to a

personal computer at a later time for analysis. Refer to Section 2.16 Importing

Saved Data into a Spreadsheet.