7 flow rate verification, Flow rate verification – Heath Consultants Hi Flow Sampler User Manual

Page 50

Advertising
background image

Maintenance

Hi Flow

Sampler

50

Instruction 0055-9017

4.7

Flow Rate Verification

This procedure verifies the Hi Flow

blower is operating at proper capacity and the flow

rate measurement is accurate. Use a calibrated flow rate meter or anemometer with a

minimum accuracy of +/-5%.

IMPORTANT: To assure accurate flow readings, conduct the following procedure in clean

air, free of any combustible gases.

1. Install the flow meter onto the Hi Flow

Sampler’s gas intake (see Figure 4-7).

2. Place the Hi Flow

Sampler into its Manual 1-Stage Mode per Section 2.14.2.2. Then

start the blower by selecting Start from the Main Screen.

3. The Flow (cfm) value as displayed on the Hi Flow

Sampler's LCD should be within

+/-5% of the measured cfm taking into account the flow meter's measurement

error and any pressure or temperature corrections.

Figure 4-7. Flow Meter Installation

4. With a fully charged battery pack, the measured flow rate should be greater than

9 cfm. If not, perform the following (see Figures 4-8 and 4-9):

• Clean the gas sample inlet screen.

• Ensure that the exhaust port is not obstructed.

• Remove the unit’s cover and ensure that the high- and low-sampling point hoses

are properly connected between the flow-measurement tube and their

corresponding solenoid valves, and that these hoses are not pinched or kinked.

• Ensure that no obstruction is present inside the flow-measurement tube.

• If none of the above solves the problem, return the unit to your nearest

Bacharach Service Center for evaluation.