Calibration

Hi Flow

Sampler

42

Instruction 0055-9017

5. Apply the appropriate level of calibration gas from the calibration equipment to the

instrument’s GAS or BACKGROUND inlet port. (In the example above, apply 2.5%
CH

4

to the BACKGROUND port.) Then press the I/O ↵ key to start the calibration

process. Observe that the gas-sampling-pump motors start and the calibration

screen appears.

6. If necessary, use the ∧∨ keys to adjust the Appl(%) reading to match the gas

concentration stamped on the calibration gas cylinder.

7. Wait until the measured gas reading stabilizes, and then press the I/O ↵ key to

calibrate the actual gas reading to that of the applied reading. The message

“Calibration Passed!” will appear at the bottom of the screen if the calibration was

successful. If, however, the calibration was not successful, the message “Calibration

Failed!” will appear. Refer to Section 5 Troubleshooting for possible causes and

remedies for this failure.

8. Disconnect the gas hose, and then wait until the measured gas reading falls to zero

percent.

9. Press the ESC key to return to the Calibration Menu Screen.

10. Repeat this procedure as necessary to calibrate both sensors at 2.5% and 100% CH