3 check calibration procedure, Check calibration procedure, Figure 3-1. calibration equipment setup – Heath Consultants Hi Flow Sampler User Manual

Page 40

Advertising
background image

Calibration

Hi Flow

Sampler

40

Instruction 0055-9017

Figure 3-1. Calibration Equipment Setup

3.3

Check Calibration Procedure

This procedure checks the calibration of both the leak-gas sensor and the background-

gas sensor using 2.5% CH4. After setting up the calibration equipment as described in

Section 3.2, proceed to check the instrument’s sensor calibration as follows.

1. Before turning ON the instrument, be sure that the calibration equipment is not

connected to the instrument’s GAS or BACKGROUND inlet port, and that the

instrument is in an area of clean air.

2. Turn ON the instrument and wait for its warm-up period to complete before

proceeding with Step 3.

3. Apply 2.5% CH

4

from the calibration equipment to the instrument’s BACKGROUND

inlet port.

4. From the Main Screen select Menu > Calibration > Verify Calibration to begin the

calibration verification process. Observe that the gas-sampling pump motors should

start.

Wait several minutes for the gas reading to stabilize. At this time the gas

concentration shown on the display should match the concentration stamped on

the calibration gas cylinder.

5. Disconnect the hose from the BACKGROUND inlet port and connect it to the GAS

inlet port.

6. Again wait several minutes for the gas reading to stabilize. At this time the gas

concentration shown on the display should match the concentration stamped on

the calibration gas cylinder.