Troubleshooting

Hi Flow

Sampler

56

Instruction 0055-9017

6

Symptom

7

Probable Cause & Remedy

Low flow rate reading with fully

charged battery.

Flow path obstructed. Check flow path and verify flow rate per Section 4.7

Slow response time.

Internal filters are dirty. Replace filters per Section 4.4.

Battery power does not last at

least 4.5 hours.

Battery pack defective. Replace.

Erratic gas readings.

a. The entire gas leak is not being captured. Use the proper attachment to

capture leak.

b. Faulty sensor(s). Calibrate both CH

4

sensors per Section 3. Replace any

sensor that fails to calibrate.

Sensor does not calibrate

properly.

a. Wrong calibration gas or insufficient flow being applied to sensor.

Ensure that the calibration setup is correct.

b. Depleted sensor. Replace sensor and recalibrate.

Calibration Failed” message

appears at bottom of display.

Calibration parameters not within programmed limits due to either

insufficient calibration gas flow, or new calibration value was greater than

50% of the old value. Check calibration setup and re-perform the

calibration procedure. Replace sensor if necessary.

Memory Full” message appears

at bottom of display.

1000 test records have already been stored in memory. Save all test

records per Section 2.15.1 and/or erase memory per Section 2.15.3.

No Records Saved” message

appears at bottom of display.

A send records command was given with no records stored in memory.

Check Background Pump

message appears at bottom of

display.

Flow to the background-gas pump has been severely reduced. Check that

the background hose is not kinked or blocked, or the internal tubing has

not become blocked (see Figure 4-10). Also check the condition of

background-gas sensor filter, replace as necessary per Section 4.4.3.

Check Leak Pump” message

appears at bottom of display.

Flow to the leak-gas pump has been severely reduced. Check that the

internal tubing has not become blocked (see Figure 4-10). Also check the

condition of leak-gas sensor filter, replace as necessary per Section 4.4.2.

Check Background Calibration

message appears at bottom of

display.

The background-gas sensor needs to be calibrated per Section 3.

Check Leak Calibration

message appears at bottom of

display.

The leak-gas sensor needs to be calibrated per Section 3.

Check Background Offset

message appears at bottom of

display.

The background-gas sensor did not zero properly during start-up. Switch

OFF instrument and restart in an area with clean air. Replace sensor if

necessary.

Check Leak Offset” message

appears at bottom of display.

The leak-gas sensor did not zero properly during start-up. Switch OFF

instrument and restart in an area with clean air. Replace sensor if

necessary.

Check Leak Rate Difference

message appears at bottom of

display.

The calculated difference between leak rates 1 and 2 was greater than

10%.