2.3

PUF Filter Adapter

For size-selective monitoring applications, it is necessary to load the appro-

priate foam filter(s) into the inlet adapter. If TSP (Total Suspended Particulate)

monitoring is required, then no foams should be fitted within the inlet

assembly.

The type of foam filter loaded into the adapter determines the size of

particulate matter being monitored by the Microdust pro and collected on the

filter. Although the PUF foam inserts have been designed as size-selective

filters to capture particles larger than a specified mean aerodynamic size, it is

also possible to weigh the PUF inserts before and after measurement to

establish both the total suspended particulate (TSP) value and the desired

size fraction.

2.3.1 Storage of PUF Filters

The PUF filters should be kept in a clean and preferably air-conditioned

environment.

2.3.2 Inserting / Removing PUF Filters

The PUF filters should be carefully inserted or removed from their respective

cassettes using clean tweezers and plastic gloves. Avoid subjecting the

filters to physical damage, creasing or folding.

Filters should be loaded into the relevant housing as shown in Figure 6.

The probe should be inserted into the Adaptor Block and correctly aligned so

that the locating spigot on the PUF Filter Holder locks it into the Probe

Measuring Chamber. When locked correctly into position, the complete

assembly is prevented from rotating or moving laterally on the probe.

2.4

Collection Filter Cassette

A Filter Cassette is provided to accommodate a 37 mm filter, (25 mm is

available on request). The desired filter type (typically of the GFA type) should

be conditioned and pre-weighed (if required) before being loaded into the

cassette. Always handle filters with care to prevent contamination or

physical damage. Never operate the system without an inlet filter being fitted

– the filter protects the pump from any ingress of dirt.

2.5

Powering Up And Using The System

Caution !

The Environmental Enclosure is completely airtight when closed. It is
necessary to unscrew and remove the pressure relief valve located
under the carrying handle before use.

Page 10 of 20

Page 11 of 20

Using the Enclosure

Using the Enclosure

P118204

P118208

Respirable Filter Holder

PM10 Filter Holder

PM2.5 Filter Holder

990009

P118206

Figure 6: Filter Housings

Pump Connection

Dia. 37 mm
Filter Holder

MICRODUST Probe

"O" Ring

Dia. 25 mm

Filter holder

(Optional)

"O" Ring

Locating Spigot

"O" Ring

PUF Filter holder

PM

/ Respirable

(Optional)

10

Dia. 37 mm to

Dia. 25 mm

Reducer

(Optional)

Adaptor

Alternativ

es

Alternativ

es

00046

Locating Spigot

PUF Filter Holder PM

2.5

PUF Filter Head Cap

Measuring Chamber

Figure 5: PUF filter installation

ENVIRONMENTAL ENCLOSURE
For MICRODUST pro - User Manual

ENVIRONMENTAL ENCLOSURE

For MICRODUST pro - User Manual