Casella CEL Environmental enclosure for Microdust User Manual

Page 6

Advertising
background image

(PM

10,

PM

2.5

or Respirable). PUF filter adaptors are described in Section 2.3

and shown in Figures 5 and 6. The PUF filter is fitted in the inlet side of the

probe mounting block and may be accessed by carefully pulling the probe

and block assembly squarely away from the case lid. DO NOT twist or

lever the probe away.

The probe should unclip from the spring clips. A nozzel on the mounting

block is a push fit into the inlet fitting on the case lid. A “purge” outlet on the

pump delivers clean air to the surface of sensitive optical components within

the measuring probe. A protective shield of clean air around the optical

components prevents possible contamination and ensures reliable, long term

4.

Check the flow if required, using a Casella -CM344 Rotameter, to be

ordered separately.

This is able to calibrate flowrates over the range from 0.5 to

5.0 litres/minute.

PUF filters are designed to operate at a flow rate of 3.5 litres/minute.

Note:

If the external power is removed from the pump, the flow setting will be lost.

3.5

Additional Pump Features

Pump ON indicator

A green LED indicates that the pump is

switched ON and is running.

Battery Limit and Shut Down Indicator

If the pump is unable to maintain the

selected flow rate due to excessive

pressure drop or an inlet blockage, the

red “Battery/Limit” LED will begin to

flash and the warning tone will bleep

rapidly. After 10 seconds the pump will

automatically shutdown.

The red “Battery/Limit” LED and a

continuous warning tone indicate that a

flow fault has occurred.

To reset the flow fault alarm, switch the

pump off by pressing both keys

simultaneously.

Low battery warning

A slowly flashing red “Battery/Limit” LED

and warning tone are used to indicate

that the battery is in a discharged

condition.

This occurs for battery terminal voltages

of approximately 4.7 volts or less and

signifies that the battery should be

re-charged.

Battery low switch off

The pump will automatically switch off

when the battery voltage falls below

approximately 4.4 volts.

This prevents a deep battery discharge

cycle and potential cell damage.

Page 6 of 20

Page 15 of 20

VORTEX Pump

Introduction

VORTEX

Pump

Green LED

Mounting
Block

Reset
Button

Inlet Tube
& Head

Red LED

Probe
Ventilation

Probe

00035

Transit

Packing

Foam Block

Purge Air

Supply

Battery

Strap

Battery

Main Power

Switch

Power

Control Unit

Microdust

Unit

"Zero"
Filter

ON/OFF
Button

Figure 2: Components and controls for the Environmental Enclosure

ENVIRONMENTAL ENCLOSURE
For MICRODUST pro - User Manual

ENVIRONMENTAL ENCLOSURE

For MICRODUST pro - User Manual

Advertising