3.6

Service Interval for Key Components

The equipment is designed and engineered for long and dependable service

but as a precision instrument it should be given the best of care. The typical

service intervals for key sampler components are as follows.

Valves

Part No 101189A

1500 hours,

Pump diaphragm

Part No 102164B

1500 hours,

Damper assembly

Part No B8351/7

1500 hours,

Eccentric 2.85 stroke

Part No 158008A

1500 hours,

Bearing

Part No BRE42

1500 hours,

Motor

Part No 104056B

3000 hours.

For rapid first line maintenance of the samplers, it is recommended that the

following stock of modular sub-assemblies also be purchased.

Diaphragm pump

Part No 151312A.

3.7

Fault Finding

SYMPTOM

POSSIBLE CAUSE

SERVICE HINT

PUMP does not run,

(green LED not lit).

BATTERY fully

discharged.

RECHARGE.

FLOW adjustment

set too low.

Turn FLOW

CONTROL up.

WIRING LOOM

disconnected at

some point.

LOCATE & re-solder.

Fault on JOINT.

REPLACE circuit

board.

PUMP does not run,

(green LED is lit).

Eccentric BEARING

SEIZURE.

REPLACE bearing.

Motor SHAFT

SEIZURE.

REPLACE motor.

Broken WIRE.

REPAIR connection.

Motor runs but PUMP IN-

OPERATIVE

GRUB SCREW

securing eccentric to

motor shaft LOOSE.

RE-TIGHTEN.

Health and Safety Executive publication MDHS 14/3. The foam filter

specifications and dimensions determine the desired aerosol size selection

characteristics and eliminate particle sizes greater than PM

10,

PM

2.5

or

Respirable (4 µm) as appropriate. The larger particles become trapped and

collect within the foam matrix, whilst all particles below these “cut-off

points” pass through the PUF filters and enter the measurement chamber,

where the real time mass concentration is established.

After passing through the Microdust pro, particulate matter is deposited on a

37 mm filter which may be used for gravimetric or chemical analysis.

Remember however, although the Environmental Enclosure has been

designed to monitor particulate fraction concentrations in a variety of

environments, it does have some operating limitations. The instrument is

not designed to operate continuously out of doors for long periods of time (in

the order of days or weeks).

Due to the light scattering principle used within the instrument, any moisture

which individual particles attract at periods of high humidity or temperatures

approaching “dewpoint” will be “seen” by the instrument as larger particles.

This biases results towards greater concentrations. Also, during these

periods, any air being drawn through the instrument which is high in humidity

may condense on the optics and distort the light beam, producing spurious

results.

With this limitation in mind, it is recommended that the Microdust pro not be

used to monitor continuously for days at a time. Any application requiring

this type of monitoring regimen, will need a particulate monitor with internal

heater, to drive off any water vapour within the incoming air stream, thus

eliminating erroneous readings.

The Casella Ambient Particulate Monitor (APM950) is considered the best

tool for this job and can run continuously and unattended for months at a

time. Please contact CASELLA CEL or your distributor for more information.

1.2

System Overview

The internal configuration of the Environmental Enclosure is shown in

Figure 2.The system incorporates a sampling pump to draw the sample air

through the inlet pipe at a user selected flow rate. PUF filters are designed to

operate at a flow rate of 3.5 litres/minute. The inlet head is designed to

prevent the ingress of insects and other large foreign objects. A dust cap is

provided to seal the inlet port on the case lid whenever the inlet tube is

removed for transit purposes.

Size selection of the sample stream is performed by passing the sample

through a PUF foam filter appropriate to the chosen sampling strategy,

Page 16 of 20

Page 5 of 20

Introduction

VORTEX Pump

ENVIRONMENTAL ENCLOSURE
For MICRODUST pro - User Manual

ENVIRONMENTAL ENCLOSURE

For MICRODUST pro - User Manual

Advertising