Introduction

1.

ENVIRONMENTAL ENCLOSURE FOR THE

MICRODUST pro

1.1

Introduction.

With the increased awareness of environmental pollution from fine

particulate and the introduction of the Air Quality legislation throughout the

world, the demand for portable real-time monitors of PM

10

and PM

2.5

has

increased. The need for real-time particulate information is also a

requirement in general industry, when looking at Total Suspended Particulate

(TSP), and at inhalable and respirable dust levels as a health issue.

CASELLA CEL have introduced the

Environmental Enclosure for exactly this

application. This accessory provides a

simple solution to fixed site monitoring

with the Microdust pro and can be used for

internal or ambient applications.

When monitoring ambient conditions, this

system should not be used under

conditions of high humidity, condensation,

rain or fog - the APM is recommended for

use under such conditions.

The IP65 case is designed to accommodate

a standard Microdust pro as well as a

sampling pump system to maintain a

precise inlet flow rate. The pump also

provides purge air to the measuring optics,

which is an important feature to maintain

long term reliability.

The Environmental Enclosure has been

designed as an accessory for existing

Microdust pro users as well as for potential

customers looking for a complete system.

The complete system would require the purchase of the Environmental

Enclosure kit PLUS the following additional items:

Microdust pro,

PUF – Polyurethane Foam Filters,

Suitable Flow Meter Kit and Sampling Filters.

The key component to this accessory is the sampling inlet which utilises PUF

size-selective filter techniques. Size-selective filters were originally developed

for operation in the Conical Inhalable Sampler (CIS) as detailed within the

Figure 1:

The Microdust pro

SYMPTOM

POSSIBLE CAUSE

SERVICE HINT

Pump runs but NO AIR

drawn

Diaphragm RUBBER

SPLIT.

REPLACE.

Pulsation damper

MEMBRANE SPLIT.

RENEW pulsation

damper.

TUBE between

pulsation

damper and valve

chest SPLIT or

DISCONNECTED.

REPLACE.

Pump runs but

CANNOT ACHIEVE

DESIRED FLOW RATE

LEAK on pump

assembly.

LOCATE & correct.

DIRT under valves or

valve DAMAGE.

CLEAN / REPLACE.

Inlet FILTER

CLOGGED.

REMOVE & CLEAN.

Flow

POTENTIOMETER

inoperative.

REPLACE circuit

board.

Flow RATE NOT

MAINTAINED through

full working shift

INCOMPLETE

CHARGE.

RE-CHARGE.

Battery BELOW

CAPACITY.

FULLY CHARGE &

check voltage.

It should be at least

12.0 V.
If below, CHECK

OUTPUT from

charger.
If output is correct,

REPLACE

BATTERY.

High current

consumption due to

MECHANICAL

DEFECT in pump.

LOCATE &

CORRECT.

Filter DISC

CLOGGED.

REPLACE.

ENVIRONMENTAL ENCLOSURE
For MICRODUST pro - User Manual

ENVIRONMENTAL ENCLOSURE

For MICRODUST pro - User Manual