Casella CEL Environmental enclosure for Microdust User Manual

Page 9

Advertising
background image

All optical measurement systems are sensitive to the presence of
moisture droplets both in the air and bonded to particulate matter. To
prevent possible measurement errors, the system should not be used in
conditions of high humidity, rainfall or condensation.

The main ON/OFF switch and Red indicator are located on the control unit

adjacent to the charger socket. When this switch is set ON, the Microdust

pro will be powered from the 12 volt lead acid battery, extending the run

time beyond the normal 20 hours (with alkaline batteries).

When the system is not being used, set this switch OFF to prevent the
12 volt lead acid battery discharging into the Microdust battery.

Switch the Microdust pro ON.

2.5.1 Checking the Instrument’s Span

The span of the Microdust pro should be checked in accordance with the

relevant User Manual; refer to Chapter 2.

The span can also be checked by inserting the calibration filter and checking

the reading. If required, adjust the span control according to the User Manual.

2.5.2 Checking Probe Zero

An inline filter is provided to simplify the task of zeroing the dust monitor.

1.

Remove the inlet head and fit the filter to the end of the inlet tube.

2.

Start the pump by pressing both buttons on the VORTEX

simultaneously.

3.

Allow the purge air to blow any residual dust from the probe.

4.

When a steady zero is achieved, set the zero calibration point of the

Microdust pro.

5.

Stop the pump by pressing both buttons on the VORTEX until the

pump stops.

6.

Remove the inlet filter and replace the inlet head.

2.5.3 Sampling Pump

Check the performance of the sampling pump and adjust the flow rate as

described in Chapter 3.

Page 12 of 20

Page 9 of 20

Using the Enclosure

Using the Enclosure

00041

Figure 4: After disconnecting

the power lead, insert a fin-

ger here to ease the unit

from the cutout

2.2

Battery Charging

Warning !

DO NOT connect the Environmental Enclosure charger directly to any
Microdust Instrument. DAMAGE to the instrument will occur.

DO NOT enable the Microdust pro internal charging circuit unless
re-chargeable batteries have been installed.

DO NOT leave the main power switch ON when the system is out of use.
The Enclosure’s 12 volt lead acid battery will discharge into the
Microdust pro
battery causing DAMAGE.

Observe the other Battery Warnings on page 3.

It is safe to leave the charger (supplied by CASELLA CEL) connected and

switched on until the unit is required for use. The charger is suitable for

indoor use only.

The Environmental Enclosure is powered by a 12 volt lead acid gel battery

which may be re-charged via the socket located on the power control unit

within the enclosure. The charger is a “split rate” type that will “fast charge”

the battery to full capacity and then switch to “trickle charge”. Plug the lead

from the charger into the socket on the power control unit and connect the

charger to the mains electricity supply.

The light on the charger will show “Red” to indicate that the battery is

charging, “Yellow” to indicate the battery has reached 80% of capacity, and

“Green” to indicate that the battery is fully charged and ready for use. Once

the battery has reached 80% capacity there is sufficient charge for at least

24 hours continuous use. When the charger is connected, the 12 volt

lead acid battery and the nickel cadmium battery in the Microdust will both

be charged at the same time.

ENVIRONMENTAL ENCLOSURE
For MICRODUST pro - User Manual

ENVIRONMENTAL ENCLOSURE

For MICRODUST pro - User Manual

Advertising