3.

THE VORTEX ASPIRATING PUMP

3.1

The Aspirating Pump

The Environmental Enclosure uses a modified version of the VORTEX

“Timer 2" personal sampler. This compact and reliable sampling pump is

designed to provide a flow rate capability of up to 4.0 litres/minute whilst

using a 37 mm GFA filter head (at up to 25 cm H

2

0 pressure drop).

Automatic flow control circuitry is used to maintain a stable sample flow rate

over varying pressure drop conditions.

The pump is powered directly from the 12 volt lead acid system battery.

3.2

Switching the Pump ON and OFF

To switch the pump on, press both keys on the VORTEX at the same time,

the buzzer will bleep twice and the green LED will be illuminated indicating

“PUMP ON”.

To switch the pump off, press both keys simultaneously until the pump

shuts down; all indicator lights should now be extinguished.

3.3

Resetting the “Elapsed Minutes” Counter

To return the count to zero, press and hold

for 3 seconds. This can be

carried out while the pump is on or off.

3.4

Adjusting The Flow Rate

The pump is factory set to a flow rate of 3.5 litres/minute. Measurement of

the sample flow should be performed by connecting a suitable rotameter to

the inlet tube.

1.

To adjust the flow rate, press and hold

until the green LED

begins to flash.

2.

Release the key.

The green LED will continue to flash for five seconds.

3.

During this time you may press

to increase the flow or

to

reduce the flow.

The flow rate will change slowly at first then more quickly the longer

the key is held down.

operation. The pump incorporates a 0.9 µm filter and adjustable bleed valve.

The valve is factory set to provide approximately 0.5 litres/minute of purge air

when the main sample flow rate is 3.5 litres/ minute.

1.3

Power Supply

A sealed, rechargeable 12 volt, lead acid gel battery is used to power the

Environmental Enclosure and provides typically in excess of 24 hours

sampling. The pump is powered directly from this battery.

Whilst the Microdust pro contains its own batteries, a small power supply

control unit is used to provide additional power from the main 12 volt battery

and extends the instrument’s normal run time.

Observe the Battery Warnings on page 3.

Three versions of the enclosure are available; select the correct one for the

local mains supply.

176091A

Enclosure for U.K. mains supply,

176092A

Enclosure for European mains supply,

176093A

Enclosure for U.S.A. mains supply.