Casella CEL Environmental enclosure for Microdust User Manual

Page 8

Advertising
background image

Page 8 of 20

Page 13 of 20

Using the Enclosure

Using the Enclosure

00040

2.

USING THE ENVIRONMENTAL ENCLOSURE

2.1

Installation

A dense foam block is used to prevent the battery from moving during

transit. Remove this transit packing block before using the Environmental

Enclosure and keep it for use the next time the Enclosure must be

transported.

Figure 3: Installing the instrument

Note also that the Microdust pro

unit is a very close fit in the foam

cutout intended for it. Install the unit

in the enclosure as follows

1.

Hold the unit over the cutout

at an angle of approximately

45

o

as shown in Figure 3.

2.

Ease the probe clip into the

slot provided for it as

shown.

3.

Lower the unit into the

cutout, ensuring that the

edge of the clip is below

the surface of the foam and

pushed firmly against the

back of the slot.

4.

Connect the power supply

cable to the charging socket

of the Microdust pro as

shown in Figure 4.

5.

When removing the

Microdust pro unit,

disconnect the power

supply from the charging

socket, then place a finger

in the slot indicated in

Figure 4 to gently lever the

unit from the foam.

The following additional steps must

be performed before the

Environmental Enclosure can be

used under operating conditions.

1.

Re-set the timer display on the pump by holding down the “RESET”

button on the VORTEX until the display re-sets to 0.

2.

Start the pump by pressing both buttons on the VORTEX simul-

taneously until the green LED comes on.

3.

Start the Microdust logger according the Microdust pro User Manual.

2.6

Starting a Sample Run

1.

Remove the protective dust cap and fit the sample inlet tube/head

to the case.

2.

Switch the Power Control unit on.

3.

Switch the Microdust pro and VORTEX pump on.

4.

Confirm the Microdust zero calibration setting using the in-line filter.

5.

Insert the PUF filter foam(s) and sample filter (as required).

6.

Check the inlet flow rate.

7.

Start the Microdust logger according to the MICRODUST pro

User Manual.

8.

Zero the pump counter if required.

9.

Leave to sample for the required run time.

ENVIRONMENTAL ENCLOSURE
For MICRODUST pro - User Manual

ENVIRONMENTAL ENCLOSURE

For MICRODUST pro - User Manual