6 wall mount sample pot – Bluelab Dosetronic Peridoser Kit User Manual

Page 14

Advertising
background image

10

Version P1.3 : 151012

Bluelab Dosetr

onic P

eridoser Kit

Bluelab Corporation Limited

Steps

1. Fix sample pot in position

Select a position on the wall that is near
the nutrient tank pump and where

all

probe cables from the Dosetronic can
reach.

2. Fit Inlet Piping

Fit a “T” junction to pipes feeding the crop
and connect supplied fl ushing ball valve.

3. Fit Outlet Piping

Connect piping to the outlet of the sample
pot and fi x into a position where it can fl ow
back into the nutrient tank.

4. Flush out plastic (swarf) and residue

Open sample pot inlet valve, allow sample
pot to fi ll and then close the valve and
allow it to drain as much as possible.


5. Adjust inlet valve

Remove lid of sample pot. Adjust fl ushing
ball valve until the water inside the sample
pot can fl ow through the outlet pipe back
to the tank.

2.6 Wall Mount Sample Pot

Conductivity

pH

Temperature

P

ipes f

eeding cr

op

To

cr
op nutrient tank

Corner

mounting screws

Flushing Ball Valve

If the Dosetronic is being installed
into a system that is already
operating, it is an option that you
could use a submersible aquarium
pump to pump water from the
nutrient tank to the sample pot.

2

Advertising