Bluelab Corporation Limited

3.10 Conductivity Calibration

Conductivity calibration is not generally
required. If you do notice that your
readings are not accurate, the fi rst thing
you should do is clean your probes.
If readings are still inaccurate the
conductivity probe can be calibrated using
a known nutrient solution of 2.77EC (see
chart for corresponding units)

1. Place conductivity probe in

conductivity standard solution.

2. Calibrate Conductivity

Select Conductivity measurement.

3. Conductivity adjustment

Adjust until actual value is 2.8EC (or
chosen unit) by using
increase / decrease keys.

2.3

-5

3

2

3

EC

CF

ppm

500

ppm

700

(TDS) (EC x 700)

2.77

27.7

1385

1939

2.8

28 1400

1960

Solution

Value

Displayed

Value

Calibration Values