Design Features

Drying surface
The PTFE coated, cast aluminum platen is resis-
tant to acidic fumes that may be released while
drying certain types of gels. A grid of vacuum
conduits on the surface allows released moisture
to be pulled away.

Vacuum port
Accepts vacuum tubing of 9-mm (3/8") ±1 mm
i.d. The port for the external vacuum source is
at the back of the instrument, at the center
of the platen.

Mains power
The mains power module houses the power cord
receptacle and one or two input fuses.

115 V~ One F 12 A, 250 V, 3 AG fuse
230 V~ Two T 6.3 A, 250 V, 5 × 20 mm

See “Care and maintenance” on page 17 for an
illustration of the mains power module.

Vacuum power cord receptacle
Connects the vacuum pump to the vacuum timer.

115 V~, 50/60 Hz. Accommodates pumps that draw
up to 5 A.
230 V~, 50/60 Hz. Accommodates pumps that draw
up to 2.5 A.