Hoefer GD2000 User Manual

Page 18

Advertising
background image

Drying thin or low-concentration polyacrylamide 
gels (
 1.5 mm) and agarose gels on paper

1

Lay a sheet of filter paper on top of the filter paper
liner and position the gel on this sheet, taking care to
avoid trapping air beneath the gel.

2

Cover the gel with thin plastic wrap. Do not leave
wrinkles in the plastic wrap.

3

Use the stiff plastic cover sheet with polyacrylamide
gels but not with agarose gels.

4

Check that the edges of all the sheets fall within the
recess of the platen.
If necessary, trim the sheets to fit within the recess.

5

Cover the stack with the silicone rubber sealing sheet.

p10

Fig 4. Stacks for drying thin or
low-concentration polyacrylamide
gels and agarose gels to filter
paper for autoradiography.

rubber sealing sheet

stiff plastic cover sheet

porous cellophane

gel

porous cellophane

filter paper liner

stainless steel screen

rubber sealing sheet

stiff plastic cover sheet*

thin plastic wrap

gel thickness

≤ 1.5 mm

filter paper

filter paper liner

stainless steel screen

*not used with agarose gels

rubber sealing sheet

porous polyethylene cover sheet

gel thickness > 1.5 mm

filter paper

filter paper liner

stainless steel screen

Advertising