Safety conditions, Emissions/immunity conditions – Measurement Computing DaqScan 2000 Series User Manual

Page 48

Advertising
background image

For clarification, terms used in some Declarations of Conformity include:

pollution degree: any addition of foreign matter, solid, liquid or gaseous (ionized

gases) that may produce a reduction of dielectric strength or surface resistivity.

Pollution Degree I has no influence on safety and implies: the equipment is at

operating temperature with non-condensing humidity conditions; no conductive particles

are permitted in the atmosphere; warm-up time is sufficient to avert any condensation

or frost; no hazardous voltages are applied until completion of the warm-up period.

Pollution Degree II implies the expectation of occasional condensation.

overvoltage (installation) category: classification with limits for transient

overvoltage, dependent on the nominal line voltage to earth. Category I implies signals

without high transient values. Category II applies to typical mains power lines with

some transients.

Safety Conditions

Users must comply with all relevant safety conditions in the user’s manual and the

Declarations of Conformity. This manual and Daq hardware make use of the following

Warning and Caution symbols: If you see either of these symbols on a product, carefully read

the related information and be alert to the possibility of personal injury.

This warning symbol is used in this manual or on the equipment to warn

of possible injury or death from electrical shock under noted conditions.

This warning/caution symbol is used to warn of possible personal injury

or equipment damage under noted conditions.

Daq products contain no user-serviceable parts; refer all service to qualified

personnel. The specific safety conditions for CE compliance vary by product; but

general safety conditions include:

The operator must observe all safety cautions and operating conditions specified in the

documentation for all hardware used.

The host computer and all connected equipment must be CE compliant.

All power must be off to the device and externally connected equipment before internal

access to the device is permitted.

Isolation voltage ratings: do not exceed documented voltage limits for power and signal

inputs. All wire insulation and terminal blocks in the system must be rated for the

isolation voltage in use. Voltages above 30 Vrms or ±60 VDC must not be applied if any

condensation has formed on the device.

Current and power use must not exceed specifications. Do not defeat fuses or other

over-current protection.


Emissions/Immunity Conditions

The specific immunity conditions for CE compliance vary by product; but general immunity

conditions include:

Cables must be shielded, braid-type with metal-shelled connectors. Input terminal

connections are to be made with shielded wire. The shield should be connected to the

chassis ground with the hardware provided.

The host computer must be properly grounded.

In low-level analog applications, some inaccuracy is to be expected when I/O leads are

exposed to RF fields or transients over 3 or 10 V/m as noted on the Declaration of

Conformity.

7-2 CE-Compliance & Noise Considerations

959195

DaqLab/2000 Series and DaqScan/2000 Series

Advertising
This manual is related to the following products: