OWON HDS Series User Manual

Page 11

Advertising
background image

OWON HDS Series Handheld DSO & DMM 2-Safety Information

applications to prevent any possible dangers.

Warning:

To avoid fire or electrical shock, please use proper power adapter. Use only the power
adapter appointed by the manufacturer and subject to approval of being used in the user’s
country.

Warning:

The two channels of the oscilloscope are non-isolated electrically. The channels should adopt
common basis during measuring. To prevent short circuits, the 2 probe ground must not be
connected to 2 different non-isolated DC level.

Warning:

The channels should adopt common basis during measuring. To prevent short circuits, the 2
probe ground must not be connected to 2 different non-isolated DC level. The diagram of the
oscilloscope ground wire connection:

It is not allowed to measure AC power when charging the oscilloscope with adapter.

Warning:

To avoid fire or electrical shock if a test tool input is connected to more 42V peak (30Vrms)
or on circuits of more than 4800VA:

Use only insulated voltage probes, test leads and adapter supplied with the test tool, or
indicated by OWON as suitable for the Oscilloscope & Multimeter.

Before use, inspect voltage probes, test leads and accessories for mechanical damage and
replace when damaged.

Remove all probes, test leads and accessories that are not in use.

Always connect the power adapter first to the AC outlet before connecting it to the
Oscilloscope & Multimeter.

Do not apply voltages that differ more than 400 V from earth ground to any input when
measuring in a CAT
environment.

Do not apply voltages that differ more than 400 V from each other to the isolated inputs
when measuring in a CAT
environment.

Do not apply input voltages above the rating of the instrument Use caution when using
1:1 test leads because the probe tip voltage will be directly transmitted to the
Oscilloscope & Multimeter.