Analog input – Super Systems DPL4000 User Manual

Page 8

Advertising
background image

Page #8 of 16___________________________________________________________________________ M4581 – DPL4000

1.

Open the menu by pressing

Open.

2.

Select Functions and press

.

3.

Select Alarm and press

.

4.

Select Quantity and press

Set to select the quantity. Since

the instrument defaults to only displaying Dew Point (Td/f), this
will be the only quantity available. Only one active quantity can
be selected at a time. To change which quantities are active, go
to Menu - Display - Quantities and units.

5.

Select the first limit and press

Set (if the alarm function is

on, turn it off). Use the arrow buttons to set the alarm level. To
select the sign for the alarm level, press the

+/- button. To

save the settings, press

OK.

6.

Select the second limit and follow the instructions in step 6. The
alarm will go off when either the upper or lower limit is
exceeded. If you will use only one limit, set the upper or lower
value to an extreme so that it will never be reached.

7.

To activate the alarm and to return to the basic display, select
Alarm On/Off and press

On. The note symbol ♫ will appear

in the upper left corner of the display.

8.

When the alarm level is reached, you can stop the alarm by
pressing

OK. To reactivate the alarm, press

Yes. To

completely stop the alarm, press

No.

NOTE

The alarm does not work when the device is turned off.

Analog Input

This feature has not been enabled on the DPL4000. It is possible to
obtain an analog output directly from the sensor. For information on
this procedure, please contact Super Systems Inc.

Advertising