E. display controls, Brightness and contrast, Instruction manual – Emerson NGA 2000 TO2 User Manual

Page 36

Advertising
background image

Instruction Manual

760008-A
September 2001

3-12 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 TO2

e. Display

Controls

To adjust the display parameters, from
the Main menu choose “Display con-
trols…” as follows:

Figure 3-14. Display Controls

NOTE:

It is possible to change the brightness
and contrast values do that the display
is no longer visible. In such case,
press the F1 key twice to change to the
multi component display and then
press the F5 key for LCDReset.

Switch automatically to “Meas-
ure” after:

This variable line allows setting of the
delay time before any selected menu
switches back to the Measure screen.
The selectable values are:

5 sec

10 sec

30 sec

1 min

5 min

30 min

Never

From the Main menu, move the cursor to the
“Display controls…” line and press the

↵ key.

The menu will display as shown below.

Brightness and Contrast:

These controls can be adjusted to accom-
modate the ambient lighting conditions. The
range of values are 20-100% for brightness
and 1-45% for contrast.

These values can be reset to the defaults
from the Multi Channel display screen (F1
from the “Measure” screen) by pressing F5
(LCDReset). It can also be reset from the
startup screen by pressing F1 (see Figure
3-8).

TO2

37.50 ppm

Measure

Status…

Channel

Lock…

MFG Data

-- Main Menu --

Analyzer basic controls (calibration) & setup…

Analyzer and I/O, expert controls & setup…

System configuration and diagnostics…

Display controls…

Time & Date: 06:34:25 February 01 2000

System tag: Fisher-Rosemount

TO2

37.50 ppm

-- Display Controls --

Brightness: 70 %

Contrast: 23 %

Switch automatically to “Measure” after: 30 s

Switch off backlight after: Never

Measure

Back…