3-16 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 TO2

Span Calibration

Depending upon the nature of your appli-
cation, it may be beneficial to verify the

span calibration of the analyzer module
every 3-4 months. The following proce-
dure illustrates how to initiate a span cali-
bration.

Figure 3-17. Span Calibration

Introduce a suitable span gas to the analyzer
and allow the reading to stabilize.

Move the cursor to the “Span gas:” line and
press the

↵ key. Enter the correct span gas

value by using the

↑ and ↓ keys to adjust the

value and the

← and → keys to select the

digit. Press the

↵ key again to save the

value.

Move the cursor to the “Calibrate to span
gas!” line and press the

↵ key. The new span

calibration will be calculated.