Signal value for 0% (100%) output, Output current, Hold output during calibration – Emerson NGA 2000 TO2 User Manual

Page 51: Instruction manual

Advertising
background image

Instruction Manual

760008-A

September 2001

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation 3-27

Model NGA 2000 TO2

Signal value for 0% (100%) out-
put

It is possible to set the signal value for 0%
output and for 100% output so as to out-
put only a portion of the entire range.

Example:

Range from 0 to 1000 ppm

0% value to be 400 ppm, 100%
value to be 700 ppm.

Analog output normally: 0V = 0
ppm, 10v = 1000 ppm

After changing the output scaling:
0V = 400 ppm, 10V = 700 ppm

Change to the “Signal value for 0% out-
put:” line and adjust the value to 400.
Then change to the “Signal value for
100% output:” line and adjust the value to
700.

NOTE:

If the measurement range is changed,
the settings done in this menu will re-
vert back to the standard values of the
range. The output values can be
changed permanently in the menu
“Begin and End of Ranges.” See sec-
tion 0.

NOTE:

The signal range of the analog output
should not be less than the smallest
range of the channel. Otherwise the

analog output may exhibit excessive
noise.

Output current

Select the desired output range in the
“Output current:” line. The options are
0…20 mA or 4…20 mA.

Hold output during calibration

Enable this option to hold the analog out-
put to the last value during calibration.

Pressing the F5 (More…) key changes to
the submenus “Output Signal if Assigned
Module Fails” and “Fine Adjustment.”

TO2

37.50 ppm

-- Analog Output Setup --

Output number: 1
Choose signal source module…
Choose signal…

Signal value for 0% output: 0.00

Signal value for 100% output: 100.00
Output current: 0…20 mA
Hold output during calibration: No
Signal name: Sample flow
Current signal value: 8.60
Source module: TO2

Measure

Back…

More…

TO2

37.50 ppm

-- Analog Output Setup --

Output number: 1
Choose signal source module…
Choose signal…
Signal value for 0% output: 0.00
Signal value for 100% output: 100.00

Output current: 0…20 mA

Hold output during calibration: No
Signal name: Sample flow
Current signal value: 8.60
Source module: TO2

Measure

Back…

More…

TO2

37.50 ppm

-- Analog Output Setup --

Output number: 1
Choose signal source module…
Choose signal…
Signal value for 0% output: 0.00
Signal value for 100% output: 100.00
Output current: 0…20 mA

Hold output during calibration: No

Signal name: Sample flow
Current signal value: 8.60
Source module: TO2

Measure

Back…

More…