Instruction Manual

760008-A

September 2001

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation 3-19

Model NGA 2000 TO2

d. Linearization

Linearization is only performed when the
sensor is changed. See section 4-3 be-
low.

e. Measurement

Display

Configuration

Figure 3-20. Measurement Display Configuration

The “Signal on mini-bargraph” lines are
used to set various status values that can
display on the four lines available on the
single component (Measure) display. Se-

lect a line and press the

↵ key. To change

the value use the

↑ and ↓ keys. The val-

ues will scroll through the allowable se-
lections as follows:

Electrolyte level

Displays “OK” or “Low.”

Range

Displays the range number from 1 to 4.

Sample flow

Displays the sample flow rate in ml/min with a bargraph.

Sensor current

Displays the sensor current in uA with a bargraph.

Temp. current

Displays the heater current in uA with a bargraph.

Sensor temp.

Displays the sensor temperature in

°C with a bargraph.

Validity

Not functional on the TO2.

Operation status

Not functional on the TO2.

Move the cursor to the “Measurement display
configuration…” line and press the

↵ key.

The “Measurement Display Configuration”
menu will be displayed.

The first two lines allow for setting the con-
centration value displayed precision and
number of digits.

The “Module identification tag:” line set the
displayed neumonic for the analyzer on all
display screens. Select the line and press the
↵ key to change the values using the ← and

→ keys to select a digit and then the ↑ and ↓
keys to change the value. The values will
scroll through all numbers and upper and
lowercase letters.

TO2

37.50 ppm

-- Analyzer Module Setup and Controls --

Calibration…
Alarms…
Ranges…
Linearization…

Measurement display configuration…

Concentration measurement…
Flow measurement…
Temperature measurement…

Measure

Back…

Status…

TO2

37.50 ppm

Measure

Back…

-- Measurement Display Configuration --

Displayed concentration digits: 6

Digits after decimal point: 2

Module identification tag: TO2

Signal on mini-bargraph – 1: Electrolyte level
Signal on mini-bargraph – 2: Temp current
Signal on mini-bargraph – 3: Range
Signal on mini-bargraph – 4: Sensor current