Instruction Manual

760008-A

September 2001

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation 3-21

Model NGA 2000 TO2

Sensor model:

Adjustable for both “Standard” and “Sta-
bel” models in the following ranges:

0-50 ppm

0-100 ppm

0-500 ppm

0-1000 ppm

0-5000 ppm

0-10000 ppm

0-5%

0-10%

0-25%

The “Stabel” sensor model contains a
special electrolyte that is less susceptible
to small concentrations of acids in the
sample stream.

Clamp low PVA to 0.02:

This parameter is used for very low oxy-
gen concentration levels where the sen-
sor current may be unstable about zero.
Normally this would be set to “No.”

ppm to mg/Nm3 conversion
factor:

The ppm to mg/Nm

3

conversion factor in

the range of 1 to 100000. The conversion
factor depends on the sample gas used.

Lower [Upper] explosion limit
(LEL) [UEL]:

The parameters are used for the CLD and
PMD modules and do not apply to the
TO2 They should both be set at 0.00%.

Temperature compensation at
span gas:

This parameter is used to set the upper
temperature limit for the sensor which can
occur at the upper range of oxygen con-
centration. The range is 10-1000 and the
default setting is 100. When the tem-
perature exceeds this level the module
goes into sleep mode.