Rosemount 3051s – Emerson 3051S User Manual

Page 26

Advertising
background image

Quick Installation Guide

00825-0100-4801, Rev HA

February 2008

Rosemount 3051S

26

European Certifications

I1

ATEX Intrinsic Safety
Certificate No.: BAS01ATEX1303X

II 1 G

EEx ia IIC T4 (T

a

= -60 °C to 70 °C)

-HART/Remote Display/Quick Connect/HART Diagnostics

1180

Special Conditions for Safe Use (X)

1. The apparatus, excluding the Types 3051 S-T and 3051 S-C

(In-line and Coplanar SuperModule Platforms respectively), is
not capable of withstanding the 500V test as defined in Clause
6.4.12 of EN 50020. This must be considered during
installation.

2. The terminal pins of the Types 3051 S-T and 3051 S-C must

be protected to IP20 minimum.

N1

ATEX Type n

Certificate No.: BAS01ATEX3304X

II 3 G

EEx nL IIC T4 (T

a

= -40 °C TO 70 °C)

Ui = 45 Vdc max
IP66

Table 2. Input Parameters

Loop / Power

Groups

U

i

= 30V

All

I

i

= 300 mA

All

P

i

= 1.0W

All

C

i

= 30nF

SuperModule

/Quick Connect

C

i

= 11.4nF

With PlantWeb or Junction Box Housing

C

i

= 0

Remote Display

L

i

= 0

All Except Remote Display

L

i

= 60 µH

Remote Display

4801_QIG_RevHA.fm Page 26 Tuesday, February 26, 2008 12:17 PM