Power and signal requirements, Installation – Ion Science Hydrosteel 7000 User Manual

Page 12

Advertising
background image

Hydrosteel 7000 MANUAL

Ion Science Ltd

Page 12 of 43

Unrivalled Detection. www.ionscience.com

Installation


Power and signal requirements
Non Intrinsically Safe (Non IS) applications
Input power

15-24 Volts 35mA

4-20 mA power

8-35 V dc. 22mA


The Hydrosteel is to be wired up as per the wiring diagram below.

























WARNING!
Non Intrinsically Safe (Non IS) applications
When the unit is installed with a non IS rated power supply the IS details on the front of the unit are not
appropriate for the system. The IS label should be covered up or blocked out. This will prevent safety
discrepancies and or mis-use in the future should the equipment be moved or the site be re-defined as a
hazardous area.

Intrinsically Safe (IS) applications
Entry parameters

Power receivers

Ui

Ii

Pi

Ci

Li

Uo

Io

Po

Units

V

mA

W

uF

mH

V

mA

mW

PS 1

30

200

1.2

0

0

PS 2

30

200

1.2

0

0

4-20mA 1

30

200

1.2

0

0

4-20mA 2

30

200

1.2

0

0

RS232

30

0

0

13

3

50

For information only see safety certificate before installation





Thermocouple

Thermocouple

junction box

1

3

E
2

SCREEN

E

3

2

+V RD

1

0V BK

Red Connector

Hydrosteel 7000

Yellow Connector

1

2

Electrical connections (Non IS)

+V RD

0V BK

SCREEN

RD

BK

BK

RD

0V BK

+V RD

SCREEN

+V

+V

0V

0V

3

E

Blue Connector

RD

BK

4-20mA 2

4-20mA 1

Control equipment

Power

+V

0V

Advertising