Fault finding and diagnostics


Fault findings
WARNING!
Ensure plant safety before starting maintenance. Certified test equipment should always be used unless a
gas clearance certificate has been issued.

The following table lists the fault outputs and their consequences. Please contact Ion Science or your local
Hydrosteel representative if you need any assistance.

Symptom

Cause

Check

No LED lights

No power,
Incorrect wiring


Fuse has blown

Ensure that there is 15 V d.c. across the power
terminals at the unit. Check the polarity; that
the positive wire goes to the + terminal.


If the voltage and polarity have been incorrectly

applied then the fuse may have blown. Note if
the fuse has blown the instrument will need to
be returned to Ion Science Ltd.

Green LED flashes
continuously. *

Insufficient power unit keeps
on resetting

Ensure that there are 15 V across -+ power
terminals at the unit.
* Note the green LED will go off for 2-3 seconds

when the pump starts.

Green with alternate

quick Flashes of the
Red LED

Fault has occurred since last

data download or last
operation. Fault is not
currently present and unit is

operating correctly

Perform a data download and look at the log of

diagnostics to identify the fault. See diagnostics
matrix below for resolve the issue.

RED LED lights

Unit has self diagnosed a
fault

Download data to find out specific diagnostic
failure. See diagnostics matrix below for resolve

the issue.

4-20 mA = 0mA

No loop power or incorrect

wiring

Ensure that there is >8 V across -+ power

terminals at the unit.
Check that positive wire goes to the + terminal.

4-20 mA = 3.5mA

Unit has self diagnosed a

fault with given
measurement. 4-20mA 1=

flux measurement ,
4-20mA 2 = pipe
temperature measurement

Download data to confirm fault and follow fault

diagnostics matrix over leaf.

RS232 connection fails Communication parameters

incorrectly set.
Incorrect modem cable.

Check red or green LED is lit
as fuse may have Blown.

Check com port settings

Check the null modem cable used

If the fuse has blown then the coms will not
operate. The unit needs to be returned to Ion

Science Ltd for servicing.