7 specifications, A. general, B. sample – Emerson 755A User Manual

Page 18: Model 755a

Advertising
background image

Instruction Manual

245364-V
May 2002

1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management

Model 755A

1-7 SPECIFICATIONS

a. General

1

Operating Range ........................... 0.00% to 100.0% oxygen
Recorder Range ............................ Selectable for 0% to 100% oxygen or for any desired span of 1%,

2%, 5%, 10%, 20% or 100% oxygen within the overall range.

Response Time ............................. (90% of fullscale) recorder output factory set for 20 seconds;

adjustable from 5 to 25 seconds.

Reproducibility (Digital Display).....

±0.01% Oxygen ±2 counts.

Ambient Temperature Limits

Maximum ............................... 49

°C (120°F) EXCEPT 38°C (100°F) for 99% to 100% oxygen.

Minimum ................................ -7

°C (20°F) EXCEPT 4°C (40°F) for 99% to 100% oxygen.

Zero and Span Drift

2

...................... Within

±1% of fullscale (±2% of fullscale for 99% to 100% range)

per 24 hours, provided that ambient temperature does not change
by more than 11.1

°C (20°F).

±2.5% of fullscale per 24 hours with ambient temperature change
over entire range.

Barometric Pressure

Compensation ....................... Oxygen readout automatically corrected to within

±1% of fullscale

for barometric pressure variations within

±3% of target value and

within

±2% of fullscale for barometric pressure variations within

±5% of target value.
The target may be set anywhere within range of -2.7 to 3.3 psig

±3

psig (-18.6 to 22.8 kPa

±21 kPa).

Exhaust vented to atmosphere.

b. Sample

Dryness ......................................... Sample dewpoint below 43

°C (110°F), sample free of entrained

liquids.

Temperature Limits

Maximum ............................... 66

°C (150°F)

Minimum ................................ 10

°C (50°F)

Operating Pressure

Maximum ............................... 69 kPa (10 psig).
Minimum ................................ -13.1 kPa (-1.9 psig)

Flow Rate

3

Maximum ............................... 500 cc/min
Minimum ................................ 50 cc/min
Recommended ...................... 250

±20 cc/min

Materials in Contact with

Sample Gas........................... 316 stainless steel, glass, titanium, Paliney No. 7, epoxy resin,

Viton-A, platinum, nickel.

1

Performance specifications based on recorder output.

2

Zero and span drift specifications based on following conditions: Operating pressure constant; ambient temperature change

from initial calibration temperature, less than 11.1 Celsius degrees (20 Fahrenheit degrees); deviation from set flow held to
within

±10% or ±20 cc/min, whichever is smaller.

3

Deviation from set flow would be held to within

±10% or ±20 cc/min, whichever is smaller. If so, zero and span drift will be

within specifications, provided that operating temperature remains constant.