1 detector/magnet assembly, 2 detector/magnet assembly wiring, Model 755a – Emerson 755A User Manual

Page 72: Instruction manual

Advertising
background image

Instruction Manual

245364-V
May 2002

6-4 Service and Maintenance

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 755A

Side View
of Connector

Keeper

Connector Pin Removed

Upper Slot

Lower Slot

Connector Pin/
Leads in Place

Improvised Pin Removal Tool, Such as a Paper Clip

Dual
Photocell

ORN

GRY

RED

BLU

BRN
YEL

Sense

Old Style Lamp

J12

10

18

1

Hole for Source Lamp

When dual photocell is
installed, the gap between
the two photocells should
be in position indicated by
this line.

Optical Bench

10

J12

1

HR2

Suspension

Heater

PUR

GRN

18

BLK

BLK

WHT
WHT

RT1

Suspension
Terminals

A. Connections to Source Lamp and Photocell B. Connections to Suspension and Heater Circuits

C. Removal and Insertion of Pin/Leads in Connector J12

Figure 6-1. Detector/Magnet Assembly

Figure 6-2. Detector/Magnet Assembly Wiring

Dual
Photocell

Connector
J12

Lamp Retaining
Set Screw

Photocell
Lock Screws (2)

Lamp Viewing
Hole

Source Lamp
Assembly

Mounting Screws (2)

Optical Bench Assembly

Detector Assembly

Magnet
Assembly

Sample Out-
let Tube

Sample Inlet
Tube

Sample Pre-Heating
Coil

Connector
Board

A. Detector/Magnet Assembly - Exploded View B. Optical Bench - Exploded View