3 calibration by pressure decrease setup, Model 755a – Emerson 755A User Manual

Page 41

Advertising
background image

Instruction Manual

245364-V

May 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9

Model 755A

Figure 3-3. Calibration by Pressure Decrease Setup

Standard Procedure: Calibration by
Pressure Decrease

1. Verify that the Pressure

Compensation Option is selected,
i.e., Jumper inserted between E1
and E3 on the Control Board.

2. Connect a compound range

pressure gauge (or mercury
manometer), pressure gauge,
pressure regulator, flowmeter,
needle valves and vacuum pump as
shown in Figure 3-3 above.

3. Open V2, set PR1 to approximately

10 psig (69 kPa) and adjust V1 for a
flow rate of 250 cc/min. into the
sample inlet line. Compound range
pressure gauge (or mercury
manometer) should indicate close to
atmospheric pressure (0.00).

4. Set TEST switch to PRESS CAL 1

and adjust CAL 1 potentiometer R98
(see Table 3-2, page 3-6, and
Figure 3-2, page 3-7) for a reading
of 00.00

±5 counts on the display.

5. Set TEST switch to NORM and note

display reading.

6. Start vacuum pump.

7. Close V2 and adjust V3 and V4 to

decrease the pressure on the
sample exhaust by about 3%. (-
12.00 inches of water; -0.4 psi; -23
mm of mercury; -3 kPa).

8. Adjust CAL 2 potentiometer R99

(see Table 3-2, page 3-6, and
Figure 3-2, page 3-7) until reading
on display is the same noted in step
5.

9. Repeat steps 3 through 7 as many

times as necessary to ensure
repeatability of compensation
calibration.

NOTE

Although not mandatory, it is
desirable, in order to obtain
maximum accuracy, that the oxygen
concentration of the gas used during
calibration be close to that of the
process stream which is to be
analyzed.

F1

Compound Range

Pressure Gauge or

Mercury Manometer

Pressure

Gauge

Vacuum

Pump

PR1

Sample

In

Sample In

755A

250

cc/min

V1

V3

10 psig

V2

V4

Advertising