B. recorder connections, C. potentiometric output, 1 electrical connections – Emerson 755A User Manual

Page 22: Model 755a, Instruction manual

Advertising
background image

Instruction Manual

245364-V
May 2002

2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 755A

N

E

U

T

H

O

T

TB1

N

E

U

T

H

O

T

TB1

Jumpers

120 VAC CONFIGURATION

240 VAC CONFIGURATION

+ - +

MA

MV

TB2

NO

COM

NC

RESET

NO

COM

NC

RESET

NO. 1

NO. 2

HOT

GND

N

E

U

T

H

O

T

TB1

COM

Jumper

GND

GND

Power Connections
(see below)

Optional Alarm Kit

mV Recorder

mA Recorder

+

-

+

-

b. Recorder

Connections

NOTE

Route recorder cable through a
separate conduit, not with power cable
or alarm output cable.

If a recorder, controller, or other output
device is used, connector it to the
analyzer via a 24-22 AWG two-conductor
shielded cable.

Route the cable through conduit to the
analyzer and into the case through the
appropriate opening shown in Installation
Drawing 642349. Connect the shield only
at the recorder or computer, if used.

Cable connections and output selection
for potentiometric and current actuated
devices are explained in below.

c. Potentiometric

Output

Insert Recorder Output Selection Jumper,
Figure 2-2 (page 2-3), in position
appropriate to the desired output; 10 mV,
100 mV, 1V or 5V.

On TB2, Figure 2-1 (page 2-2), connect
leads of shielded recorder cable to "MV+"
AND "COM" terminals.

Connect free end of output cable to
appropriate terminals of recorder or other
potentiometric device.

For device with a span of 0 to 10mV, 0 to
100mV, 0 to 1V, or 0 to 5V, connect cable
directly to input terminals of the device,
making sure polarity is correct.

For device with intermediate span, i.e.,
between the specified values, connect
cable to device via a suitable external
voltage divider, as shown in Figure 2-3
(page 2-3).

Figure 2-1. Electrical Connections