Instruction Manual

245364-V
May 2002

3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 755A

TEST Switch

ZERO Adjust

Rosemount Analytical

M o d e l 7 5 5 A

O x y g e n A n a l y z e r

PRESS CAL1 PRESS CAL 2

NORM REC OFFSET

6 Digit LCD Display

SPAN Adjust

ZERO
SPAN

CONTROL

FUNCTION

Digital Display (LCD)

Readout of sample oxygen content (0.00% to 100.00%) or selected test function,
depending on position of TEST switch.

TEST Switch (S1)

Selects variable desired for readout on digital display.

Switch Position

Readout on Digital Display

Designation

1

Oxygen content of sample (0.00% to 100.00%)

NORM

2

Pressure/Voltage CAL 1

CAL 1

3

Pressure/Voltage CAL 2

CAL 2

4

Zero suppression applied to analog output circuit
(0.00 to 10.00 VDC)

ZERO SUPPRESSION

ZERO Control (R13)

Used to establish downscale calibration point on digital display or recorder chart.
With suitable downscale standard gas flowing through the analyzer, the ZERO
Control is adjusted for appropriate reading on display.

SPAN Control (R16)

Used to establish upscale calibration point on digital display or recorder chart. With
suitable upscale standard gas flowing through the analyzer, the SPAN Control is
adjusted for appropriate reading on display or recorder.

Figure 3-1. Model 755A Front Panel Controls