Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-13

Model 755A

b. Selection of Deadband

The desired deadband may be selected
with the appropriate trimming
potentiometer:

ALARM 1 = R73

ALARM 2 = R78

For any setpoint, deadband is adjustable
from 1% of fullscale (counterclockwise
limit) to 20% of fullscale (clockwise limit).
Deadband is essentially symmetrical with
respect to setpoint.

Figure 3-4. Schematic Circuit of Alarm Relay Assembly

+15V

-15V

ALARM 1

COMMAND

ALARM 2

COMMAND

J5

1

2

4

6

14

CR1

13

K1

1

5

9

12

8

NO

COM

NC

RESET

ALARM 1

CR2

13

K2

1

5

9

12

8

NO

COM

NC

RESET

ALARM 2

14

2. CR1 AND CR2 ARE ANY 600V, 1A DIODE.
1. RELAYS SHOWN IN ENERGIZED POSITION.

NOTES: