6-6 Service and Maintenance

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 755A

If the lamp assembly removed from the
instrument has four wires, the 633689
Connector Board requires modification.
Continue to step 10.

10. Refer to Figure 6-1 on page 6-4. Remove

the two screws holding the Connector
Board to the magnet assembly. Carefully
remove Connector Board.

11. Place Connector Board on a clean

working surface, with solder side (no
components) up.

12. Per Figure 6-4 below, add straps or solder

bridges at the two points shown.

NOTE

If the Connector Board cannot be
satisfactorily modified, a modified 633689
Connector Board may be ordered from the
factory. See Section 7.

13. Reassemble detector, etc., in reverse

order of disassembly.

14. Refer to Figure 6-5 below. The red line

on the lamp must be aligned with the
retaining set screw. Insert lamp into
mounting hole until it extends 1/4 inch.
Tighten set screw.

15. Realign the photocell per Section 1-1a on

page 6-7.

Figure 6-4. Modification of 633689 Connector Board for Compatibility with Replacement Lamp

Figure 6-5. Lamp Alignment

1/4"

Red Mark for
Alignment

Set Screw

Solder Side of Board (Backside)

F3

HR1

Add Straps or Solder Bridges