1-13 MicroCEM SPECIFICATIONS

a. Analyzer

Power ............................................ Universal Power Supply 85 – 264 VAC, 50 – 60 Hz, + 10%, 500

Watts Maximum at Start Up. 250 Watts Nominal

Microprocessor.............................. Intel Pentium processor, 266MHz, 64MB RAM, PC/104 architec-

ture, Windows NT embedded Platform

Pocket PC...................................... 133MHz, 21 bit Hitachi SH3 processor, 32MB RAM 16MB ROM,

240 X 320 pixels LCD, Riuch color, backlit, Wireless LAN optional

Detectors//Number ........................ NDIR, Paramagnetic, Electrochemical, Chemiluminescense// Up to

three in one analyzer

Mounting........................................ Wall Mount
Area Classification......................... General Purpose / NEMA 4X (IP65) Fiberglass Enclosure
Ambient Range

Temperature ......................... -30

° to 50° Celsius.

Relative Humidity................... 5 to 99%

Inputs/Outputs

Digital..................................... RS-485 Serial Port. (Multi-Drop Network)

RS-232 Serial Port.
LAN, Ethernet 10/100-BaseT
Modem.

Connectivity Protocols ........... HTML (Web Browser) – Status, file transfer Modem/Webrowser

TCP/IP..Modbus (In Process)
Foundation Fieldbus (In Process)

Analog Current Outputs......... 3 Isolated 4-20 mA DC, 500 ohms Max Load (O

2

, CO, NO

X

)

Analog Inputs:MW, Fuel Flow

Digital OutputsTrouble Alarm, Sample Pump on/off, Drain Pump
on/off, Purge on/off, Calibrate on/off – 110VAC @ 1amp Dry Con-
tact

O

2

Limit Exceed, CO Limit Exceed, NOx Limit Exceed, O

2

Low

Range, CO Low Range, NO

X

Low Range. - TTL: 5 VDC Max Cur-

rent 20 mA

Digital Inputs.......................... Process on/off, Initiate Auto Calibration

Instrument Weight ......................... 62 lbs. Typical
Dimensions.................................... 24“ x 20“ x 12“ (HxWxD)
Ranges .......................................... O

2

: 0 –25%

CO: 0 –100ppm Selectable to 1000ppm
NO

X

: 0 – 10ppm Selectable to 1000ppm

Sample Temperature..................... 0

° C to 55° C

Sample Flow Rate ......................... .5 to 1.5 liters/min
Warm Up Time .............................. Max 25 minutes @ low ambient temperatures