3-14 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model MicroCEM

General Setting

Description

O2SensorLowLimit

This sets the O2 % low limit, below which a limit alarm will occur.

O2SensorHighLimit

This sets the O2 % high limit, above which a limit alarm will occur.

COSensorLowLimit

This sets the CO low limit, below which a limit alarm will occur.

COSensorHighLimit

This sets the CO ppm high limit, above which a limit alarm will occur.

NOXSensorLowLimit

This sets the NOx low limit, below which a limit alarm will occur.

NOXSensorHighLimit

This sets the NOx ppm high limit, above which a limit alarm will occur.

CCEM Serial Number

This setting is used to record the MicroCEM Serial Number.

Diluent Percent

Percent O2 used in Diluent correction.

MinimumOnCurrent

This is the minimum current that must be measured by a current analog input for a

heater/cooler/fan to be considered on. This is used for heater/cooler/fan failure de-

tection.

MaximumOffCurrent

This is the maximum current that must be measured by a current analog input for a

heater/cooler/fan to be considered off. This is used for heater/cooler/fan failure de-

tection.

NDIR Sync Low

This indicates which NDIR reading is made when the sync goes from high to low:

Reference (R) or Sample Gas (S).

+24V Low Limit

If the +24V measurement is below this level it will cause an alarm.

+24V High Limit

If the +24V measurement is above this level it will cause an alarm.

ValveOnTime

For calibration – The amount of time to wait after turning on a valve.

O2EmissionLimit
NOXEmissionLimit
COEmissionLimit
O2LowRange
COLowRange
NOXLowRange
O2HighRange
COHighRange
NOXHighRange

These limits are the default values. On the Pocket PC these values will be used

when one of the “Defaults” buttons are pressed.

SamplePumpOn

1 = Sample pump is used. 0 = Sample pump not used.

LogsDir

Directory where log files are saved (C:\uCEM\Logs).

MaxWarmUpTime

Maximum time allowed for the uCEM to warm-up when it is started up. If all tem-

perature zones are not within allowed range within this period of time, the uCEM will

shutdown all heaters, coolers and processes.

StateFile

The full path to the file which maintains the state of the uCEM.

(C:\uCEM\mcem.state)

ServerIP

The IP address to use when listening for incoming connections.

ServerPort

The IP Port to use when listening for incoming connections.

ServerTimeout

100

AlarmsPersistFile

The full path to the file which maintains the current state of the alarms.

(C:\uCEM\alarms.state)

LogoffTimeout

Auto Log-off time in minutes

UserListPersistFile

The full path to the file that maintains the user list (C:\uCEM\mcem.userlist)

Table 3-4. Factory Settings - General