e. Calibration Log File Format

The log file contains data in a flat, AS-
CII, CSV file. The following are the

fields of the file, in order of occurrence.
The log file size will be about 178 bytes
per entry. 3 months of data logs will
require about 16000 bytes (based on
Calibration performed every 24 hours).

Name

Description

Example

Date/Time

Month-day-year Hours:Minutes:Seconds

3-7-2001 10:24:57

Zero Time

Time that Zero span started, Hours:Minutes:Seconds

10:25:30

Mid Time

Time That Mid span started, Hours:Minutes:Seconds

10:27:30

Span Time

Time that span started, Hours:Minutes:Seconds

10:28:30

Purge Time

Time that the final purge started,
Hours:Minutes:Seconds

10:30:30

Finish Time

Time that the final purge finishes

10:31:00

O2 Measured Zero

Measured percent O2 for Zero phase of calibration

0.0

O2 Expected Zero

Expected percent O2 for Zero phase of calibration

0.0

O2 Zero Drift

Percent drift of O2 zero calibration

0.0

O2 Measured Mid Span

Measured percent O2 for Mid span phase of calibration

10.1

O2 Expected Mid Span

Expected percent O2 for Mid span phase of calibration

10.0

O2 Mid Drift

Percent drift of O2 mid calibration.

-0.4

O2 Measured Span

Measured percent O2 for Span phase of calibration

20.2

O2 Expected Span

Expected percent O2 for Span phase of calibration

20.3

O2 Span Drift

Percent drift of O2 span calibration

0.4

CO Measured Zero

1

CO Expected Zero

0

CO Zero Drift

0.3

CO Measured Mid Span

23

CO Expected Mid Span

24

CO Mid Span Drift

-0.3

CO Measured Span

45

CO Expected Span

45

CO Span Drift

0

NOx Measured Zero

Measured ppm NOx for zero phase of calibration

15

NOx Expected Zero

Expected ppm NOx for zero phase of calibration

15

NOx Zero Drift

0

NOx Measured Mid Span

Measured ppm NOx for mid span phase of calibration

30

NOx Expected Mid Span

Measured ppm NOx for mid span phase of calibration

30

NOx Mid Span Drift

0

NOx Measured span

Measured ppm NOx for span phase of calibration

59

NOx Expected span

Measured ppm NOx for span phase of calibration

59

NOx Span Drift

0

Table 3-8. Calibration Log File Format