2-8 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model MicroCEM

a. AC Power

Connect AC power through a 10A circuit
breaker that is to be located close to the
MicroCEM. The circuit breaker will pro-
vide over current protection as well as a
means of disconnecting the power.

Maximum power requirements will be 380
watts, with most applications requiring
less than this amount.

b. Circular Connector Assembly Instruc-

tions

Wire Stripping

Strip insulation from end of wire to be
crimped. Do not cut or damage wire
strands. Refer to table for proper strip-
ping dimensions.

Wire Size

Dim. A

22O or 22M*

.125 (3.18)

20

.188 (4.77)

16

.188 (4.77)

12

.188 (4.77)

*Inactive. Not recommended for new
design, replacement only.

Contact Crimping

1. Insert stripped wire into contact crimp

pot. Wire must be visible through in-
spection hole.

2. Using correct crimp tool and locator,

cycle the tool once to be sure the in-
dentors are open. Insert contact and
wire into locator. Squeeze tool han-
dles firmly and completely to insure
a proper crimp. The tool will not re-
lease unless the crimp indentors in
the tool head have been fully actu-
ated.

3. Release crimped contact and wire

from tool. Be certain the wire is visi-
ble through inspection hole in contact.

A

Visual Inspection Hole