5-6 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Model MicroCEM

position of the O-ring and window in the
upper compartment.

The new photodiode should be slowly in-
serted into the housing while gradually
rotating the body. This allows the O-ring
to properly seat. Continue replacing
screws, washers, thermistors, etc., with
the thicker shim (washer) on the opposite
side of the socket from the thermistor.

Replace the lower section of the housing,
then the bottom cover, insulator and
bracket with the shoulder washers and
screws.

Re-tighten the screws in the Reaction
Chamber (upper section). Replace the top
cap and its screws.

To reinstall in the Analyzer Module, re-
verse the procedure for removal as indi-
cated above.

Figure 5-4. Chemiluminescense Detector Assembly

Photodiode

Ozone

Detector Mounting

Bracket

Photodiode Socket

Assembly

Thermistor

Assembly

Reaction

Chamber

Sapphire
Window

Exhaust

Sample