Hoshizaki STACKABLE CRESCENT CUBER KM-1301SAH-E User Manual

Page 53

Advertising
background image

53

4) Close the low-side valve and high-side valve on the gauge manifold.
5) Disconnect the gauge manifold hose from the vacuum pump and attach it to a

refrigerant service cylinder. Remember to loosen the connection and purge the air
from the hose. For air-cooled and water-cooled models, see the nameplate for the
required refrigerant charge. For remote air-cooled model, see the rating label inside the
icemaker. Hoshizaki recommends only virgin refrigerant or reclaimed refrigerant which
meets ARI Standard 700 (latest edition) be used.

6) A liquid charge is required when charging an R-404A system (to prevent fractionation).

Place the service cylinder on the scales; if the service cylinder is not equipped with
a dip tube, invert the service cylinder, then place it on the scales. Open the high-side
valve on the gauge manifold.

7) Allow the system to charge with liquid until the proper charge weight is met.
8) If necessary, add any remaining charge to the system through the low-side.

CAUTION! To prevent compressor damage, use a throttling valve or liquid
dispensing device to add the remaining liquid charge through the low-side
service valve with the unit running.

9) Close the high and low-side service valves. Close the high and low-side gauge manifold

valves, then disconnect the gauge manifold hoses.

10) Cap the service valves to prevent a possible leak.

Advertising
This manual is related to the following products: