GE GEK-106465A User Manual

Page 42

Advertising
background image

ACCEPTANCE TESTS

GEK-106465A

CIO Remote CAN Digital I/O Module

41

6.6. COMMUNICATION

WITH THE MASTER UNIT

1. Connect the master unit to the CIO. Verify that the BUS LED blinks indicating communication between both

units.

2. Verify, using the master unit Setup program, that the CIO boards setting options are available. For example,

using an F650 as master unit, verify in EnerVista F650 Setup, that the "Actual > Inputs/Outputs > I/O
Board Status > Board H Status
" and "Actual > Inputs/Outputs > I/O Board Status>Board J Status"
appear as active.

6.7.

INPUTS AND OUTPUTS

During all tests, the screw on the rear of the relay must be grounded.

6.7.1. DIGITAL

INPUTS

During this test, the user will determine the activation/deactivation points for every input in the relay for the set voltage
value of 30 Volts.

Verify that the error does not exceed +/- 10% (+10% on activation, -10% on deactivation)

Default board settings for the input test are as follows:

Settings/Control Elements/Inputs Outputs/Board X

I/O Board Type 1

Voltage Threshold A_01

30 V

Voltage Threshold B_01

40 V

Debounce Time A_01

15 ms

Debounce Time B_01

15 ms

Input Type 01 00 (CC1)

POSITIVE

... ...

Input Type 01 15 (CC16)

POSITIVE

The inputs test is completed by groups of 8 inputs, as this type of board has 2 groups of 8 inputs with the same
common. For the first 8 inputs, the voltage threshold setting is determined by Voltage Threshold A. For the next 8
inputs, the setting is Voltage Threshold B.

Inputs (or contact converters, CC1 – CC15) must also be set to POSITIVE.

I/O Board Type 2

Voltage Threshold A_01

30 V

Voltage Threshold B_01

40 V

Debounce Time A_01

5 ms

Debounce Time B_01

5 ms

Input Type 01 00 (CC1)

POSITIVE

... ...

Input Type 01 07 (CC8)

POSITIVE

Advertising