Electrical installation – Invacare IH6300ADH User Manual

Page 21

Advertising
background image

SECTION 4—INSTALLATION

Part No 1096926

21

Supine Bath Tub

Electrical Installation

DANGER

The National Electric Code requires the use of a ground fault circuit interrupter on all

therapeutic pools and tubs (hydrotherapeutic tanks). Failure to follow this requirement

could result in serious or fatal injury.
ELECTRICAL SHOCK HAZARD. All disassembly and maintenance of the tub MUST

be done by a qualified technician, certified electrician or plumber. Disconnect electrical

supply or turn OFF circuit breaker before performing any maintenance to the tub.

ƽ WARNING

Be certain ALL electrical work is performed by a qualified electrician and is in

compliance with local electrical codes.
All electrical conduit, fittings and wire to unit are to be supplied and installed by a

professional electrician in accordance with the National electrical code as a minimum

requirement, including the supply and installation of a Ground Fault Circuit Interrupter

(GFCI).
Before installing and connecting the tub, make certain that the electrical power source

is shut off to avoid electrical shock.

NOTE: For this procedure, refer to FIGURE 4.4.
1. Install a 115V electrical hook‐up per local 

electrical codes.

2. For Model IH6300, perform the 

following:
A. Remove the screws that attach the 

rear access panel to the tub.

B. Route electrical line through electrical 

line routing hole in tub frame.

C. Connect ground, neutral and hot 

wires from wall directly to motor 
through the electrical line routing 
hole.

DANGER

Model IH6300ADH - Electric shock haz-

ard. Use only with hospital grade GFCI

receptacle. Failure to do so may result in

an electrical shock and/or serious or fatal

injury.

3. For Model IH6300ADH ‐ Plug hospital 

grade plug into pre‐installed GFCI 
protected hospital grade receptacle 
(Detail “B” of FIGURE 4.4).

I

FIGURE 4.4 Electrical Installation

Rear Access

(Panel Removed)

Electrical Line

Routing Hole

Screw (1 of 4)

Hospital Grade Plug

Motor

DETAIL “A” - MODEL IH6300

DETAIL “B” - MODEL IH6300ADH